CLEANING AND DISINFECTION PROTOCOL
 
 
ABRASIVE DISC

Remove abrasive disc from handpiece. Clean disc after each procedure using warm water and mild detergent. Disinfect with quaternary ammonia
or hydrogen peroxide after each use. If blood is present, soak in bleach for 20 to 40 minutes (or refer to the bleach manufacturer’s specific directions for appropriate contact time) or clean with neutral ph enzymatic detergent. Dry thoroughly with a clean lint-free cloth, then store in its container and return container to carrying case.

FINGER GUARD

Clean guard using an alcohol, quaternary ammonia or hydrogen peroxide moistened lint-free cloth or equivalent wipe. 
Do not use corrosive agents to clean/disinfect. If blood is present, soak in bleach for 20 to 40 minutes (or refer to the bleach manufacturer’s specific directions for appropriate contact time or clean with neutral ph enzymatic detergent. Dry thoroughly with a clean lint-free cloth; return to carrying case.

CLEAR SHIELD

Clean clear plastic shield using an alcohol, quaternary ammonia or hydrogen peroxide moistened lint-free cloth or equivalent wipe. Do not use corrosive agents to clean/disinfect. 

HANDPIECE

Clean handpiece using an alcohol, quaternary ammonia or hydrogen peroxide moistened lint-free cloth or equivalent wipe. Do not use corrosive or abrasive agents to clean/disinfect. Return handpiece to carrying case. WARNING: DO NOT IMMERSE HANDPIECE IN LIQUID.
 
 
#250 RING SPREADER

Clean spreader using an alcohol, quaternary ammonia or hydrogen peroxide moistened lint-free cloth or equivalent wipe. 
Do not use corrosive agents to clean/disinfect. If blood is present, soak in bleach for 20 to 40 minutes (or refer to the bleach manufacturer’s specific directions for appropriate contact time or clean with neutral ph enzymatic detergent. Dry thoroughly with a clean lint-free cloth; return to carrying case.
 


 